§ 874.4180. Eustachian tube balloon dilation system.

304 words·~1 min read·/us/cfr/t21/s§ 874.4180·

A research copy — for the controlling text, always check the official state or federal source. Not legal advice.

(a)Identification. A Eustachian tube balloon dilation system is a prescription device that includes a flexible catheter attached to an inflatable balloon. The system is intended for use in dilating the cartilaginous portion of the Eustachian tube for treating persistent Eustachian tube dysfunction.
(b)Classification. Class II (special controls). The special controls for this device are:
(1)Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be evaluated:
(i)Mechanical testing, including tensile and flexural testing of catheter joints and materials.
(ii)Durability testing, including fatigue and burst pressure testing of the balloon materials and components.
(iii)Inflation and deflation characterization testing, including time and pressure measurements, and leak testing of the balloon.
(iv)Verification testing of safety features built into the device must be performed, including the characterization of catheter geometries and distal tip insertion limitation mechanisms.
(2)Simulated use testing in a clinically relevant model must demonstrate the reliability of the device to remain mechanically functional throughout the anticipated conditions of use, and validate that the design features limit access to only the cartilaginous portion of the Eustachian tube.
(3)The patient-contacting components of the device must be demonstrated to be biocompatible.
(4)Performance data must demonstrate the sterility of the device.
(5)Performance data must support shelf life by demonstrating continued sterility of the device, package integrity, and device functionality over the identified shelf life.
(6)Training must include simulated use on cadavers to ensure users can follow the instructions for use to allow safe use of the device.
(7)Labeling must include:
(i)Detailed instructions for use.
(ii)A detailed summary of the device technical parameters, including maximum allowed inflation pressure, allowable catheter geometries, and available balloon sizes.
(iii)A shelf life. [81 FR 73041, Oct. 24, 2016]
Connections1 cite this
Cited by 1 section
Citation graph
cites case law
§ 874.4180
Eustachian tube balloon dilation system.
Fed. Reg.×1
Cites 0Cited by 1 across 1 source
★   the supreme law of the land   ★
Don't Tread on Me
E Pluribus Unum — out of many, one

"If you don't know your rights, you don't have any."

Marginalia · a citizen's law index
A research desk, not legal advice. Always read the cited source before relying on a summary.
Questions or an issue? support@self-law.org
disclaimerMarginalia is a research index, not a law firm. Nothing on this site is legal, tax, or financial advice and no attorney–client relationship is formed by using it. Statutes, regulations, and case law change; summaries, search results, AI output, and member posts may be incomplete, out of date, or wrong. Any interpretation drawn from material on this site should be validated by a licensed attorney in your jurisdiction before you act on it.